DESCRIPTION
SCR
ELEKTRONIKS have developed Patient / Earth Leakage Current Testers
to measure the Leakage Current in medical appliances. The Patient
/ Earth Leakage Current Tester Model - PELCT-MS isdesigned to carry
out Leakage Current measurement in 3 modes, namely :
1. Earth / Enclosure Leakage Current Test.
2. Patient Leakage Current.
3. Patient Input Current.
Earth / Enclosure Leakage Current Test is useful to measure Leakage
Current through Earth terminal or through metal enclosure. This
is applicable to equipment like X-Ray Machines and other Medical
Equipments.
Patient Leakage Current and Patient Input Current Tests are useful
for ECG Monitors. The Internal Circuits are so designed that, only
with the help of different switch positions the tests can be carried
out.
TECHNICAL FEATURES / SPECIFICATION
01. Digital Display for Leakage Current Range : 3 1/2 Digit.
02. Current measurement Ranges : 200 Micro Amps., 2000
Micro Amps. and 20 mili Amps.
03. Current Measuring Accuracy : +/- 2% + 5 micro Amps.
04. Current Trip : If the current is more than the full
range the supply is disconnected.
05. Trip Indication on LED.
06. Polarity Switch is provided to change the polarity
of the output supply.
07. Standard 3 pin socket for easy connection of Equipment
Under Test.
08. Selection of Tests :
1) Earth / Enclosure Leakage Current Test.
2) Patient Leakage Current Test.
3) Patient Input Current Test.
09. Optional Facility to Carry out test for equipment with
3 Phase Supply Input.
10. Facility to connect five probes at a time.
11. Facility to select individual probe one by one or all
probes at a time.
12. Facility to test X Ray - 2 Modes as follows :
1) One model working on 3 phase power supply with open
ended terminal with connections .
2) Second model working on 15 A power supply socket from
mains.
13. Input Supply 230 V +/- 10% AC 50 HZ.
|