Phantom Load(Test
Setup) for Single Phase Energy Meter
The panel comprise of the following :
1. Power source 0 to 300 V AC, 0 to 50 A with Power Factor adjustment
facility from 0.1 Lag to 0.1 Lead through unity with seven steps.
2. Digital Voltmeter : - 0 to 300 V AC.
3. Digital Ammeter : - 0 to 1.000 reading with CT.
4. Current Transformer having ranges 2A, 5A, 20A, 50A.
5. Digital Power Factor Meter : - 0.1 Lag - Unity - 0.1 Lead.
6. Mains ON MCB : - This is TPN MCB of 4 A Rating provided for
switch " ON " or " OFF " the panel.
7. CT Range Switch : - This is provided for selecting the CT Range.
8. TEST ON / OFF Push Buttons : - This is provided to ON / OFF
the output of the panel.
9. Adjust Voltage Variac : - This is Single Phase / 2 A Variac
provided for adjusting the output voltage from 0 to 300 V.
10.Adjust Current Variac : - This is a Single Phase / 2 A Variac
provided for adjusting the load current from 0 to the range selected
by switch.
11.Set Power Factor Switch : - For adjusting the P. F. one switch
with seven steps is provided. With this coarse adjustment of P.
F. can be done.
12.Adjust P. F. Fine Variac : - This is a Single Phase / 2 A
Variac provided to adjust P. F. After selecting P. F. with switch,
we have fine adjustment with this Variac.
13.Output Terminals : - Two terminals of voltage & two terminals
for current are provided at the front panel. With these terminals
we can connect the `Meter under Test' to the panel.
Phantom Load(Test
Setup) for Three Phase Energy Meter
For 3 phase Energy Meter testing,
3 such systems are housed in a single panel
TECHNICAL SPECIFICATIONS FOR THREE PHASE POWER SOURCE / PHANTOM
LOAD UNIT
Three
Phase Power Source for Testing of 40 Energy Meters will comprise
of Voltage & Current Source for 40 Nos. meters with following
Specifications for EACH PHASE :
a. Input supply - 3 phase 50 Hz, 415 V 10 %.
b. Outputs - I. Voltage Source, II.Current Source.
I. Voltage Source : Adjustable 0 - 300 Volt/600VA with
Static phase shifting arrangement to obtain 90 Deg. lead through
0 to 90 Deg. lag with respect to Current Source. Facility to switch
to 70% and 120% of set voltage will be provided for creep test.
II. Current Source : 0 - 100 Amps. adjustable in steps
with switch and fine by variac; Switching arrangement will be
made to get 100%, 400%, 500%, 600% and 800% of basic setting.
Also arrangement will be made to select 0.4 %, 5%, 10%, 20%, and
100% of setting for carrying out Starting Current Test.
c. Digital Ammeter 3 & 1/2 digit Class 1 with CT range
selector.
d. Digital Voltmeter 3 & 1/2 digit 0 - 300 V class
1.
e. Power factor meter to read phase angle between 90 Deg
lag-Unity-90 Deg lead.
f. Line frequency Indicator 40 hz to 55 hz
g. Outputs will be provided on Insulated terminals of
proper sizes.