SCR ELEKTRONIKS Have Developed A System For
Calibration Of Static Energy Meters. This Is A PC Based
System Which Facilitates To Test Number Of Meters Say 5,
10, 15, 20, 40 Or 80 At A Time Resulting In Speedy Testing
Of Meters.
The Test Panel Provides Facilities To Carryout
Following Tests As Per IS : 13779 : 1999
1. Accuracy Check At UPF, 0.5 Lag P.F., 0.8 Lead P.F.At
Different Current Values.
2. Starting Current Test ( At 0.4% Of Basic Current )
3. Creep Test ( At 70% & 120% Rated Voltage)
4. Repeat Accuracy Test.
|
The System Will Comprise Of -
1. Power Source
2. Error Display Units Depending On Number Of Stations.
3. Photo Electronic Sensors
4. Fabricated Panel For Connecting The Meters To The System
5. Computer System
6. Software
7. Reference Standard Meter
Description :
1.Power Source : It Comprises Of Voltage & Current
Source With Following Specifications -
A. Input Supply - 3 Phase 50 Hz, 415 V +/- 10 %.
B. Outputs - I. Voltage Source II. Current Source
I. Voltage Source : Adjustable 0- 300 Volt
/ 1KVA With Static Phase Shifting Arrangement To Obtain 90 Deg
Lead And 90 Deg Lag With Respect To Current Source. Facility
To Switch To 70 % And 120 % Of Set Voltage Is Provided For Creep
Test.
II. Current Source : 0 - 200 Amps Adjustable In Steps With Switch
And Fine By Variac. Switching Arrangement Is Made To Get 100
%, 400 %, 500 %, 600 % And 800 % Of Basic Setting. Also Arrangement
Is Made To Select 0.4 %, 5 %, 10 %, 20 %, And 100 % Of Setting.
C. Digital Ammeter 3 & 1/2 Digit Class 1
With CT Range Selector.
D. Digital Voltmeter 3 & 1/2 Digit 0 - 300
V Class 1.
E. Power Factor Meter To Read Phase Angle Between
90 Deg Lag-Unity-90 Deg Lead.
F. Line Frequency Indicator 40 Hz To 55 Hz.
G. Outputs Are Provided On Insulated Terminals Of Proper Sizes.
2. Error Display Unit : It Is Microcontroller
Based 6 Digit Display Unit To Indicate +/- % Error (-100.00 To +100.00).
It Takes Pulse Input From Standard Meter And LED Input From MUT
& Provides Required Data To Computer. It Also Provides Indication
Of PASS Or FAIL In Case The Error Is Within Or Out Of Limit Respectively.
3. Photo Electronic Sensor : For Each Station
Sensing Head Is Provided To Detect The Blinking Of The LED On Meter.
4. Fabricated Panel : Aesthetically Fabricated
Panel To House Display Units And Arrangement To Connect Meters Is
Provided. The Arrangement For Connection Of The Meters Will Be Simple
For Operation.
5. Computer System : It Can Be A Standard
Computer System Having Minimum One ISA Slot On Its Motherboard.
SCR Make ADDON Card Is Inserted In It For Communication With Display
Units. Also One Printer, May Be Bar Code Printer Is Required For
Printing Sr.No. Etc. On Meter.
6. Software : User Friendly Software Is Provided
Which Enables Operator To Select Different Test Parameters And Carryout
The Test. It Also Generates Different Reports. Also It Provides
Facility To Print Bar Code.
7. Reference Standard Meter : Reference Standard Meter Such
As Radian Make Type RM 10, RM 12 For Single Phase Meters & MTE
Make Type SWS 1.3 For 3 Phase Meters Are Used To Obtain Pulses Which
Are Proportional To The Energy Consumed. These Pulses Are Given
To Error Display Unit For Error Calculation.
|